Sensitivity of non-ophthalmic ultrasound probe in the localization of metallic intra-ocular foreign bodies / Jerome Mary S. Sarmiento
Subject(s): Abstract: Industrialization carries an increased stress to include ocular trauma. Yet in a not so highly urbanized locale, very sophisticated and expensive diagnostic instruments that help diagnosis and specific localization of intraocular foreign bodies are not currently available. A previous study made at Cebu Doctors Hospital Ultrasound Center using a non-opththalmic (a 5 Mhz rectal probe) in the detection of metallic intraocular foreign bodies using porcine eyes revealed a 100 sensitivity. In the study, the authors would like to determine the sensitivity using the same prove in the localization of metallic intraocular foreign bodies randomly placed in four quadrants in porcine eyes. A total of 75 eyes were used inthe study. Fifteen subjects eyes were used per quadrant plus the control group. The ultrasound probe was noted to be most sensitive at the antero-superior quadrant (93) (p0.05) and least sensitive at the postero-inferior quadrant (73) (p0.01). In the group with no foreign bodies (control), sensitivity was 73 (p0.01).| Cover image | Item type | Current library | Home library | Collection | Shelving location | Call number | Materials specified | Vol info | URL | Copy number | Status | Notes | Date due | Barcode | Item holds | Item hold queue priority | Course reserves | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Journal Article
|
DOH Central Library Electronic Resource Section | J000267 (Browse shelf(Opens below)) | Available | D0001J000267 |
Browsing DOH Central Library shelves,Shelving location: Electronic Resource Section Close shelf browser (Hides shelf browser)
Industrialization carries an increased stress to include ocular trauma. Yet in a not so highly urbanized locale, very sophisticated and expensive diagnostic instruments that help diagnosis and specific localization of intraocular foreign bodies are not currently available. A previous study made at Cebu Doctors Hospital Ultrasound Center using a non-opththalmic (a 5 Mhz rectal probe) in the detection of metallic intraocular foreign bodies using porcine eyes revealed a 100
sensitivity. In the study, the authors would like to determine the sensitivity using the same prove in the localization of metallic intraocular foreign bodies randomly placed in four quadrants in porcine eyes. A total of 75 eyes were used inthe study. Fifteen subjects eyes were used per quadrant plus the control group. The ultrasound probe was noted to be most sensitive at the antero-superior quadrant (93) (p0.05) and least sensitive at the postero-inferior quadrant (73) (p0.01). In the group with no foreign bodies (control), sensitivity was 73 (p0.01).
Cebu Doctors' Proceedings, 1994 10 (2) pages 127-31
There are no comments on this title.